취급품목
Gateway to Electrochemistry
Gateway to Electrochemistry
SYSTEM | |
Potentiostat | Yes |
Galvanostat | Yes |
Zero Resistance Ammeter | Yes |
Floating (Isolated from Earth Ground) | Yes |
Cell Connections | 2, 3, 4, 5, or 6 (includes Aux channel) |
Maximum Current | ±600 mA |
Current Ranges | 11 (60 pA - 600 mA) |
Current Ranges (including internal gain) | 13 |
Minimum Current Resolution | 20 aA |
Maximum Applied Potential | ±11 V |
Rise Time | <250 ns |
Minimum Time Base | 3.333 μs |
Noise and Ripple | <2 μV rms |
Weight | 3 kg |
Dimensions | 9 (W) x 19 (H) x 27 (D) cm |
CONTROL AMPLIFIER | |
Compliance Voltage | ±22 V |
Output Current | >±600 mA |
Speed Settings | 5 |
EIS MEASUREMENT | |
EIS | 10 μHz - 5 MHz |
EIS Accuracy | See Accuracy Contour Plot |
Voltage AC Amplitude | 3 V maximum |
Current AC Amplitude | 600 mA maximum |
ELECTROMETER | |
Input Impedance | >1014 Ω || <0.2 pF |
Input Current (typical) | <10 pA |
Bandwidth | >15 MHz at -3 dB |
CMRR | 65 dB (1 MHz) |
POTENTIAL | |
Applied Accuracy | ±1 mV ±0.2% of setting |
Applied Resolution | 200 μV, 50 μV, 12.5 μV/bit |
Measured Accuracy | ±1 mV ±0.2% of reading |
Measured Resolution | 1 μV, 10 μV, 100 μV, 400 μV/bit |
CURRENT | |
Applied/Measured Accuracy | ±10 pA ± 0.05% of range ±0.2% of value (600 mA - 6 nA) or 0.75% of value (600 pA) or 1.5% of value (60 pA) |
Applied/Measured Resolution | 0.003% full-scale/bit |
Bandwidth | >10 MHz (600 mA – 600 μA) >0.15 MHz (6 μA)) |